  • 18/08/2025

    ok

    Comments
    Show Original
  • 31/10/2025

    Received the units that were ordered._When I tested them they did not perform well._Below 100MHz the readings are very unstable._The instability appears to be caused by either the micro, the display or the combination of the two as the output voltage from the detector is very stable._IFR 2031A signal generator set at -10dBm, power meter calibrated to -10dBm at 900MHz, the results were as follows:_At 1.0MHz readings fluctuated between -4.4 to -39.0dBm_At 2.0MHz readings fluctuated between -4.6 to -30.3dBm_At 5.0MHz readings fluctuated between -2.1 to -26.0dBm_At 10.0MHz readings fluctuated between 1.7 to -30.6dBm_At 20.0MHz readings fluctuated between -5.6 to -15.0dBm_At 50.0MHz readings fluctuated between -10.4 to -11.3dBm_At 100.0MHz reading was -10.8dBm_At 200.0MHz reading was -10.6dBm_At 500.0MHz reading was -10.4dBm_At 900.0MHz reading was -10.0dBm_At 1.0GHz reading was -10.0dBm_Both units appear to behave the same.